Convention Paper

Acoustic Evaluation of Prosthetic Cardiac Valve Performance

Authors: Kingsley, Benedict; Waxler, Edward B.; Green, Peter

AES Convention 41 · Paper 807 · October 1971

Abstract

A new non-invasive technique, octave band analysis has been utilized for measuring objectively the sound level energy in dB and cutoff frequency in Hz of prosthetic valve sounds in 100 patients. Frequency profiles (dB vs. frequency) have been plotted for normally functional Starr-Edwards silastic and metal ball valves, Cross-Jones and Beall disc valves and Bjork flap valves. For example, a normal aortic metal ball prosthesis showed a frequency cutoff at 16,000 Hz. The same type of aortic prosthesis with massive thrombosis showed a cutoff frequency of 500 Hz. Four cracked and deformed silastic aortic balls demonstrated considerable reduction in dB energy level as well as cutoff frequency of 250 Hz. This technique appears to be very useful for the detection of prosthetic valve malfunction.

Details

Published in
AES Convention 41
AES Convention
41
Paper number
807
Publication date
October 6, 1971
Affiliation
Hahnemann Medical College & Hospital, Division of Cardiology, Philadelphia, PA (See document for exact affiliation information.)
Type
Convention Paper